4 Operation
732 IC Detector / 733 IC Separation Center
120
• Identification "ZEROREF"
Reference value for the auto-zero function (measured value which is
compensated).
• Identification "OVERLOAD"
The measured value is 50…80% outside the selected full-scale
range.
• Identification "BACKGROUND"
The measured value is more than 80% outside the selected full-scale
range. The absolute conductivity (background) is outputted in place
of the auto-zero value.
Output of individual measured values
• For the output of individual measured values, ">CONFIG/print
meas.value = immed." must be selected (see section 4.4.2). Each
time the <PRINT> key is pressed, a measured value will be out-
putted via the RS232 interface.
• Example:
Individual measured value output with print header output
(with date and time), run time, measured value
>CONFIG/printer
id.1
id.2
print header: always
date&time: on
send to: Seiko
>CONFIG/print meas.value
print crit.: immed.
date&time: off
732 IC Detector 01104 732.0012 Print header
date 1995-10-09 time 13:24:58 Date and time
12.3 138.9 uS/cm Run time, measured value
==============
732 IC Detector 01104 732.0012 Print header
date 1995-10-09 time 13:25:37 Date and time
139.1 uS/cm ZEROREF Auto-zero reference value
13.0 +0.0 uS/cm Run time, auto-zero value
==============