4 Operation
732 IC Detector / 733 IC Separation Center
112
>PROGRAM/edit. XXX.X min
valve B: ---
---, fill, inject
Switch injection valve B
This inquiry appears only with "SETUP/peripherals/operation
with 733 = on" and with the 733.0020 IC Separation Center with
two injection valves.
fill Switching of injection valve B to "FILL" posi-
tion (equivalent to the <FILL> key, see sec-
tion 4.6.1).
inject Switching of injection valve B to "INJECT"
position (equivalent to the <INJECT> key,
see section 4.6.2).
>PROGRAM/edit. XXX.X min
suppressor: ---
---, step
Switch suppressor module
This inquiry appears only with "SETUP/peripherals/operation
with 733 = on" and with the 733.0X30 IC Separation Center with
suppressor module.
step Switching of the suppressor module to the
next position (equivalent to the key function
<STEP>, see section 4.6.1).
>PROGRAM/edit. XXX.X min
remote TTTTTTTT
T, 0, 1
<CLEAR> sets all values to T
Set remote output lines 1…8
Sets the remote output lines 1…8 (from left to right) to
the following values:
0 Line off, inactive (open).
1 Line on, active (0 V).
T Leave line in current status
(entry of "T" with numeric key <.>).
You will find further details of the remote output lines in
section 6.2.
>PROGRAM/edit. XXX.X min
pump R/S: ---
---, on, off
Start/stop pump drive
This inquiry appears only with "SETUP/peripherals/operation
with 709 = on".
on Switch on pump drive of the 709 IC Pump
(equivalent to the <PUMP R/S> key, see
section 4.6.6).
off Switch off 709 IC Pump Drive (equivalent to
the <PUMP R/S> key, see section 4.6.6).
>PROGRAM/edit. XXX.X min
flow: --- mL/min
---, 0.015 mL/min
Reset flow rate
This inquiry appears only with "SETUP/peripherals/operation
with 709 = on".
The flow rate of the 709 IC Pump is set to the inputted
value (for further details of the "flow" parameter, see
section 4.5.1).