4.5 System state window
IC Net 2.3
39
E295
Storage of memory values failed (repeat
last action; if error reappears, call Metrohm
service).
E296
Instrument stopped (restart instrument; if
error reappears, call Metrohm service).
E297
Storage of remote line values failed (repeat
last action; if error reappears, call Metrohm
service).
E298
Storage of flow correction value failed (re-
peat last action; if error reappears, call
Metrohm service).
E299
Storage of break time values failed (repeat
last action; if error reappears, call Metrohm
service).
E300
High-pressure pump faulty (restart pump; if
error reappears, call Metrohm service).
E301
Injection valve blocked (check injection
valve; if error reappears, call Metrohm ser-
vice).
E302
Suppressor module blocked (check sup-
pressor; if error reappears, call Metrohm
service).
E303
Storage of maintenance information failed
(repeat last action; if error reappears, call
Metrohm service).