Metrohm IC Net 2.3 Bedienungsanleitung Seite 108

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 389
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 107
6 Devices
IC Net 2.3
98
E1, E2, E3 Working potentials.
Range: -2 ... +2 V.
t1, t2, t3 Time duration to apply the working poten-
tial.
Range: 0.01 ... 2 s
Sample time Measuring time at the end of working po-
tential
E1.
Selection:
20, 40, 60, 80, 100 ms
Oven Switching on/off column oven and setting
operating temperature.
Range:
15 ... 60 °C
Range Selection of measuring range:
pA 10 ... 5000 pA.
nA 0.1 ... 50 nA.
µA 0.01 ... 5 µA.
Full scale The full scale range (operating range) sets
the desired sensitivity for the analog output.
The possible values of the full scale range
depend on the preset measuring Range.
Offset Offset of the zero point for the analog out-
put signal.
Selection: -50 ... +50 % (in steps of 10 %)
Seitenansicht 107
1 2 ... 103 104 105 106 107 108 109 110 111 112 113 ... 388 389

Kommentare zu diesen Handbüchern

Keine Kommentare