6.28 844 UV/VIS Compact IC
IC Net 2.3
257
Startup with hardware The activated devices will be started with
Startup hardware
or at the start of a determi-
nation.
UV lamp If this option is switched off the UV lamp will
not be switched on at Startup hardware or at
the start of a determination.
VIS lamp If this option is switched off the VIS lamp
will not be switched on at Startup hardware
or at the start of a determination.
Column oven If this option is switched off the column
heating will not be switched on at Startup
hardware
or at the start of a determination.
Peristaltic pump If this option is switched off the peristaltic
pump will not be switched on at Startup
hardware
or at the start of a determination.
Reference channel Reference channel.
On If this option is switched on the absorption
of the measuring and reference channels
will be shown as a comparative value.
Wavelength, nm Sets the wavelength of the reference chan-
nel.
Bunching, nm Sets bunching for the wavelength defined
above.
Links
The
Links tab of the 844 UV/VIS Compact IC is used for COM port
selection and settings (details see section 5.2.4 Links).