6.12 Metrohm 709 Solvent Delivery Unit
IC Net 2.3
137
Program table
Number Row number in the program table (read-
only).
Time Time at which the flow and/or gradient is
changed.
Entry range: 0.0 ... 99999 min
Flow Overall flow of pump control system in
mL/min.
Entry range: 0.01 ... 20.00 mL/min (depending
on number of pumps)
%B Individual flow rate for pump B in percent of
total flow.
Entry range: 0 ... 100 %
%C Individual flow rate for pump C in percent of
total flow.
Entry range:
0 ... 100 %
%D Individual flow rate for pump D in percent of
total flow.
Entry range:
0 ... 100 %
Enabled Enable time program start (a disabled pro-
gram is not started).
Show flow column Show Flow column in the program table.
<Add> Add a new line to the time program.
<Delete> Delete the selected line from the time pro-
gram.
<Start> Start all pump drives.