7.2 Passport
IC Net 2.2
143
7.2.1 General
Tab General of the passport with general description of the
method and determination.
Ident User defined identifier (title) for the chroma-
togram to be displayed in the title bar of the
chromatogram window and in the Chroma-
togram
open window.
Duration Duration of the chromatogram in min. You
can modify this value when the chroma-
togram is running.
METHOD Path and file name of the method used for
data acquisition (read-only).
DATA Path and file name of the current chroma-
togram (read-only).
Date/time Date and time of the chromatogram start
(read-only).
Last update Date and time of the last chromatogram
modification and saving (read-only).
Calibration level If the current run is used for calibration, the
calibration level is a positive integer, other-
wise it equals 0. You can modify this value
when the chromatogram is running.
User Full name of the current user (read-only).
Detector Name of the detector. It can be entered
with Method setup / Measure.
Run Number of the current run starting from the
very first one. All runs are automatically
numbered by the system (read-only).