6 Devices
IC Net 2.2
98
Manual
The Manual tab of the 753 Suppressor Module window for manual
operation is only available for a connected system.
Pump
<On> Start pump drive.
<Off> Stop pump drive.
Suppressor
<Step> Rotate suppressor to the next position. The
time since the last step action is displayed
in the field beside the <Step> button.
Start pump with startup hardware
Automatic start of pump drive with Startup
hardware
.
Time program
On the
Program tab of the 753 Suppressor Module window a user-
defined time program can be entered. This program is started
automatically as defined in the
Start mode window (see 4.4.3) ei-
ther at the moment the determination is started (Start with determi-
nation
) or at the moment the sample is injected (Start with inject).