8 How to ...?
797 VA Computrace – Software
228
Dosino 2 is used to add Brightener standard solution. That should
also be indicated on the
Substances tab of the EDIT WORKING
METHOD PARAMETERS
window (here with CVS):
Sample table
List the samples in turn on the Sample table, and connect them to
the according method (note the method parameters described
above):
Start determination.
Brightener Analysis with 838 Advanced Sample Processor and "LAT"
"LAT" should only be used, if the difference between Q(Intercept
solution + Bath solution) and Q(only Intercept solution) is too small
for "MLAT".
Following installations and settings are recommended for the au-
tomated Brightener determination with the 838 Advanced Sample
Processor and Calibration technique "LAT":