3.2 Remote control commands
795 KFT Titrino
82
.MeasPt.X"165" Time of the MPList in s
.MeasPt.Y"3.654" Volume of the MPList in mL
.MeasPt.Z1"6.34" Measured value of the MPList, format depends on the
measured quantity
.MeasPt.Z2 without meaning
.EP.X"1.234" Volume coordinate of the EP
.EP.Y"5.34" Measured value coordinate of the EP
3.2.2.72. Info.ActualInfo.Oven.HeatTime read only
Info.ActualInfo.Oven.SampleTemp read only
Info.ActualInfo.Oven.LowTemp read only
Info.ActualInfo.Oven.HighTemp read only
Info.ActualInfo.Oven.GasFlow read only
Info.ActualInfo.Oven.UnitFlow read only
$Q sends the current values from a connected KF Oven. If no Oven is con-
nected, the values are empty.
.HeatTime: Heating time of sample in s.
.SampleTemp: Nominal sample temperature in °C.
.LowTemp: Lowest temperature during the sample heating time in °C.
.HighTemp: Highest temperature during the sample heating time in °C.
.GasFlow: Average gas flow during sample heating time.
.UnitFlow: Unit of gas flow.
3.2.2.73. Info.ActualInfo.Display.L1 up to 32 ASCII characters
Info.ActualInfo.Display.L8 up to 32 ASCII characters
Info.ActualInfo.Display.DelAll $G
Lines of the display. The display can be written to from the computer. Pro-
ceed as follows:
1. Lock the display, see 3.2.2.86.
2. Delete the whole display (.DelAll).
3. For writing onto the display, the standard character set will be used.
4. Unlock the display, see 3.2.2.86.
5. Delete the whole display (.DelAll).
6. Send a value to nod &Config.Aux.ResDisplay (see 3.2.2.46) to refresh the
display.
$Q sends the contents of the corresponding display line.
3.2.2.74. Info.ActualInfo.Comport.Number read only
$Q sends the comport number of the Titrino where the PC is connected.
3.2.2.75. Info.Assembly.CycleTime read only
Info.Assembly.ExV read only
Inquiries regarding basic variables of the assembly: Cycle time in s, volume of
the active Exchange Unit in mL.