7.2 Reports
104
780/781 pH/Ion Meter, Manual
The settings shown for the instrument configuration correspond to the
standard settings in the normal operating condition of the pH/Ion Meter
after the configuration memory has been initialized (see Section 8.5) fol-
lowed by conversion to English as the dialog language.
7.2.7 Parameter report
in order to document the parameter settings for the current method you
can produce a parameter report:
'pa
781 pH/Ion Meter 01104 5.781.0020
date 2004-06-03 12:08:36
PARAMETER pH
>measuring parameters
electr.id
drift 0.050 /min
method ********
delta measurement: OFF
stirrer: OFF
>calibration parameters
drift 0.5 mV/min
report: OFF
cal.interval OFF h
no.of buffers 2
buffer type: Metrohm
low lim.slope 95.00 %
up lim.slope 103.0 %
low lim.pH(0) 6.40
up lim.pH(0) 8.00
offset Uoff state: OFF
>limits pH
state: OFF
>limits T
state: OFF
>plot parameters
left mar.: 0.0
right mar.: 14.0
left marg.T 20 °C
right marg.T 30 °C
>preselections
req.ident: OFF
>electrode test
electrode type: standard
report: OFF
--------------
The standard parameters following a memory initialization in the current
mode are shown; in this case the pH mode with a connected tempera-
ture sensor (see Section 8.5). Such parameters can be stored as a
method (see Section 6.1).
PARAM
REPORT
9
+