4.4 Optimum working conditions
■■■■■■■■■■■■■■■■■■■■■■
40
■■■■■■■■
851 Titrando
4.4 Optimum working conditions
4.4.1 General
When a coulometer cell that has been well dried-out beforehand is put
into operation with a generator electrode without diaphragm, the basic
drift will be reached within approx. 30 minutes. It is recommended that
the coulometer cell be repeatedly and carefully shaken during this time.
For generator electrodes with diaphragm, you should expect a preparation
time of approx. 2 hours.
To obtain precise determinations of amounts of water smaller than
100 µg, it may also be of advantage to condition the coulometer cell over-
night before using it.
If the Titrando is switched off for extended periods with the coulometer
cell filled, it will take some time for it to become conditioned again after it
is switched back on.
With continuous use, the Titrando should not be switched off overnight.
4.4.2 Drift
A constant drift in the range of 4 µg/min is all right. Lower values, how-
ever, are quite possible. Higher but stable values will still produce good
results, because the drift can be compensated.
A constantly high drift can be caused by water-containing deposits in
inaccessible parts of the coulometer cell. In these cases, shaking the cell
can reduce the value. Ensure that there are no drops above the liquid level
in the coulometer cell.
If you are working with a generator electrode with diaphragm, shake the
cell only hard enough that the catholyte and anolyte do not mix with one
another. If the drift remains too high for a prolonged time, even after
shaking the cell, then the electrolyte solutions should be replaced. The
catholyte should be replaced once per week.
A wet catholyte can be another reason for the excessively high drift. The
wet catholyte can be dried with a KF single-component reagent.
When you work with a Karl Fischer oven, a drift of 10 µg/min is all right.
The drift depends on the gas flow (the smaller the gas flow, the lower the
drift) and on the humidity of the surroundings.