2.1 Instrument setup 711 Liquino, Instructions for use 4 2 Installation 2.1 Instrument setup Packaging The 711 Liquino is supplied in packagin
4.6 CONT mode - preparation of solutions 711 Liquino, Instructions for use 94 4.6 CONT mode - preparation of solutions The CONT mode (content)
4. Detailed description 711 Liquino, Instructions for use 95 Dosing procedure The necessary weight of the original substance is calculated when th
4.6 CONT mode - preparation of solutions 711 Liquino, Instructions for use 96 4.6.3 CONT mode parameters PARAMB All settings in the parameter m
4. Detailed description 711 Liquino, Instructions for use 97 >CONT par amet er cont ent st d. 1. 0 Content of standard solution 0.
4.6 CONT mode - preparation of solutions 711 Liquino, Instructions for use 98 >CONT pr esel ect i ons >>CONT pr ompt s Submenu for def
4. Detailed description 711 Liquino, Instructions for use 99 4.6.4 Example of a report - CONT mode Preparation of solutions and dilutions Parame
4.7 PIP mode - precise pipetting and diluting 711 Liquino, Instructions for use 100 4.7 PIP mode - precise pipetting and diluting The PIP mode
4. Detailed description 711 Liquino, Instructions for use 101 Dosing units If you only want to pipet without any dilutions then select a 2 mL or 5
4.7 PIP mode - precise pipetting and diluting 711 Liquino, Instructions for use 102 Pipetting with the pipetting equipment In principle the proce
4. Detailed description 711 Liquino, Instructions for use 103 4.7.1 PIP mode parameters PARAMB All settings in the parameter menu form a method
2. Installation 711 Liquino, Instructions for use 5 • Insert fuse holder Insert fuse holder and press it until it clicks into place. 230 2.2 S
4.7 PIP mode - precise pipetting and diluting 711 Liquino, Instructions for use 104 par amet er s >PI P r at es Dosing and filling rate sett
4. Detailed description 711 Liquino, Instructions for use 105 4.7.2 Example of a report - PIP mode Pipetting with dilution Parameter Report 71
4.8 SMPL mode - fully automatic sampling 711 Liquino, Instructions for use 106 4.8 SMPL mode - fully automatic sampling The SMPL mode (sampling)
4. Detailed description 711 Liquino, Instructions for use 107 • Start determination: a start signal is transferred to an externally connect-ed in
4.8 SMPL mode - fully automatic sampling 711 Liquino, Instructions for use 108 PREP — preparation is important All the Dosinos needed for the sam
4. Detailed description 711 Liquino, Instructions for use 109 Example: remote connection: Titroprocessor 726 Liquino (under Parameter >SMPL c
4.8 SMPL mode - fully automatic sampling 711 Liquino, Instructions for use 110 4.8.1 Sampling process in detail The SMPL mode procedure is laid
4. Detailed description 711 Liquino, Instructions for use 111 Rinse Dosino rinseDosinorepeatrinsingemptyDosino • The Dosino aspirates the samp
4.8 SMPL mode - fully automatic sampling 711 Liquino, Instructions for use 112 Add auxiliary solution addaux. solution • With further Dosin
4. Detailed description 711 Liquino, Instructions for use 113 4.8.2 SMPL mode parameters PARAMB All settings in the parameter menu form a metho
2.2 Safety information 711 Liquino, Instructions for use 6 cate the source of the fault with the aid of the diagnosis functions. If the cause of
4.8 SMPL mode - fully automatic sampling 711 Liquino, Instructions for use 114 >SMPL sampl i ng oper at i on sampl e r i nse r ep. 3 Nu
4. Detailed description 711 Liquino, Instructions for use 115 >SMPL meas. cel l r i nsi ng r i nsi ng vol ume 50. 0 mL Solvent volume for
4.8 SMPL mode - fully automatic sampling 711 Liquino, Instructions for use 116 >SMPL dosi ng uni t s >>aux. sol ut . AFTER Submenu fo
4. Detailed description 711 Liquino, Instructions for use 117 >>ext . devi ce cont r ol awai t acknowl edge: of f Type of acknowledgm
4.8 SMPL mode - fully automatic sampling 711 Liquino, Instructions for use 118 4.8.3 Example of report - SMPL mode Parameter Report 711 Li qui
4. Detailed description 711 Liquino, Instructions for use 119 Result Report 711 Li qui no 0220 711. 0011 dat e 1998- 10- 15 t
4.9 GLP mode - validation 711 Liquino, Instructions for use 120 4.9 GLP mode - validation With the 711 Liquino Dosinos or dosing units can be co
4. Detailed description 711 Liquino, Instructions for use 121 (cover top of weighing chamber). Place the Erlenmeyer flask on the weigh-ing pan. •
4.9 GLP mode - validation 711 Liquino, Instructions for use 122 ceptance can be previously set in the parameter menu under '>GLP Op-tions
4. Detailed description 711 Liquino, Instructions for use 123 4.9.1 GLP mode parameters PARAMB All settings in the parameter menu form a method
2. Installation 711 Liquino, Instructions for use 7 2.3 Setup 2.3.1 Connecting the keyboard The keyboard is connected to the keyboard socket p
4.9 GLP mode - validation 711 Liquino, Instructions for use 124 >GLP opt i ons fill rate ma x mL / mi n Filling rate 0.001…166.66,
4. Detailed description 711 Liquino, Instructions for use 125 4.9.2 Example of report - GLP mode Validation of dosing units Parameter Report 71
4.9 GLP mode - validation 711 Liquino, Instructions for use 126 Detail Report 711 Liquino 0220 711.0011 date 1998-10-03 time
4. Detailed description 711 Liquino, Instructions for use 127 Curve showing the absolute error ← limit value according to DIN ← li
4.10 Storing and loading methods 711 Liquino, Instructions for use 128 4.10 Storing and loading methods The parameters of the different instrume
4. Detailed description 711 Liquino, Instructions for use 129 met hods >del et e met hod Delete method >del et e met hod met hod:
4.10 Storing and loading methods 711 Liquino, Instructions for use 130 Submenus: met hod di r >r ecal l met hod Load method from card The
4. Detailed description 711 Liquino, Instructions for use 131 met hod di r >del et e di r ect or y Delete directory >del et e di r ect or
4.10 Storing and loading methods 711 Liquino, Instructions for use 132 met hod di r >change bat t er y Enter date of next battery change for
5. The 711 Liquino in synthesis laboratories 711 Liquino, Instructions for use 133 5 The 711 Liquino in synthesis laboratories The modes which ar
2.4 Memory card 711 Liquino, Instructions for use 8 2.3.4 Installing the pipetting attachment The pipetting attachment available as an option i
5.1 The advantages of the 711 Liquino 711 Liquino, Instructions for use 134 process until the temperature is again within the limits, interruptio
5. The 711 Liquino in synthesis laboratories 711 Liquino, Instructions for use 135 If 'intelligent' thermostats with a suitable interfac
5.2 Please note … 711 Liquino, Instructions for use 136 chloromethane, trinitrophenol-2,3,6, vinyl acetate, vinyl chloride, vinyli-dene chloride
5. The 711 Liquino in synthesis laboratories 711 Liquino, Instructions for use 137 Two useful and simple methods for auxiliary tasks which are oft
5.2 Please note … 711 Liquino, Instructions for use 138 conf i gur at i on … >dosi ng uni t s 1 I D RI NSE l en
6. Appendix 711 Liquino, Instructions for use 139 6 Appendix 6.1 Dosing units and Dosinos Four 700 Dosinos can be connected to the 711 Liquino a
6.1 Dosing units and Dosinos 711 Liquino, Instructions for use 140 700 Dosino 10 ml10 ml Remove dosing drive The dosing drive mounted on the d
6. Appendix 711 Liquino, Instructions for use 141 Port 0 – is used for venting the storage vessel and is usually fitted with an absorber tube (f
6.1 Dosing units and Dosinos 711 Liquino, Instructions for use 142 Emptying — but automatic A comfortable function is also available for compl
6. Appendix 711 Liquino, Instructions for use 143 6.1.2 Defining dosing units In the configuration menu (press <CONFIG>) under '<Do
2. Installation 711 Liquino, Instructions for use 9 2.5 Connections Metrohm automated systems with the 711 Liquino RS 232RemoteDosimatspH Meter
6.2 Error messages 711 Liquino, Instructions for use 144 6.2 Error messages * 50mL cyl . not possi bl e In the PIP mode 50 mL dosing units
6. Appendix 711 Liquino, Instructions for use 145 DOSxx: wr ong dos. t ype The connected dosing device is not a 700 Dosino. Switch off the Liq
6.2 Error messages 711 Liquino, Instructions for use 146 Error messages relating to data transmission: Note: If neither a printer nor a computer
6. Appendix 711 Liquino, Instructions for use 147 * * * Rat e t o l ow. Adj ust par amet er ! * * * The necessary dosing rate required to ful
6.3 Service and maintenance 711 Liquino, Instructions for use 148 6.3 Service and maintenance 6.3.1 Service The 711 Liquino should be serviced
6. Appendix 711 Liquino, Instructions for use 149 6.4 Validation / GLP GLP (Good Laboratory Practice) requires, among other things, the regular c
6.5 Diagnosis 711 Liquino, Instructions for use 150 6.5 Diagnosis 6.5.1 General The 711 Liquino is a very precise and reliable control instru
6. Appendix 711 Liquino, Instructions for use 151 6.5.2 Overview Access to the diagnosis main menu is described in section 6.5.3. By pressing
6.5 Diagnosis 711 Liquino, Instructions for use 152 6.5.4 Check working memory (RAM) This diagnosis step carries out a non-destructive test of
6. Appendix 711 Liquino, Instructions for use 153 6.5.6 Check display contrast This diagnosis step allows the contrast setting variability func
2.5 Connections 711 Liquino, Instructions for use 10 2.5.1 Remote connections The remote control of instruments within a Metrohm automated syste
6.5 Diagnosis 711 Liquino, Instructions for use 154 6.5.8 Check remote interfaces This diagnosis step allows the functioning of all outputs an
6. Appendix 711 Liquino, Instructions for use 155 6.5.10 Check external bus This diagnosis step allows the functioning of those parts of the ha
6.5 Diagnosis 711 Liquino, Instructions for use 156 Provided that the memory card remains inserted during the whole test proce-dure and that the
6. Appendix 711 Liquino, Instructions for use 157 6.5.13 Check memory card interface This diagnosis step allows the functioning of the internal
6.5 Diagnosis 711 Liquino, Instructions for use 158 • Press <ENTER>. After pressing the <ENTER> key the interface Dos.1 will be a
6. Appendix 711 Liquino, Instructions for use 159 6.5.16 Check software reset This diagnosis step allows a warm start (software-controlled inst
6.6 Initialize data memory 711 Liquino, Instructions for use 160 6.6 Initialize data memory This diagnosis step allows standard values for the
6. Appendix 711 Liquino, Instructions for use 161 By pressing the <SELECT> key the submenus are selected in sequence. Access to the indivi
6.7 Technical specifications 711 Liquino, Instructions for use 162 6.7 Technical specifications Short description Compact, multi functional do
6. Appendix 711 Liquino, Instructions for use 163 RS232-Interface Socket D-subminiature 9-pin Connection for PC, printer or balance programma
2. Installation 711 Liquino, Instructions for use 11 Connection cables Only Metrohm cables should be used to connect the 711 Liquino to other inst
6.8 Warranty and certificates 711 Liquino, Instructions for use 164 6.8 Warranty and certificates 6.8.1 Warranty The warranty regarding our pro
6. Appendix 711 Liquino, Instructions for use 165 6.9 EC Declaration of conformity The METROHM AG company, Herisau, Switzerland hereby certifies
6.9 EC Declaration of conformity 711 Liquino, Instructions for use 166 6.9.1 Certificate of Conformity and System Validation This is to certify
6. Appendix 711 Liquino, Instructions for use 167 6.10 Accessories 711 Liquino 2.711.0010 includes the following accessories: Mains cable with
6.10 Accessories 711 Liquino, Instructions for use 168 Reagent bottles Reagent bottle with GL 45 thread, 1000 mL, amber glass 6.1608.023 Reagent
6. Appendix 711 Liquino, Instructions for use 169 Pipetting equipment Pipetting handle 6.1562.040 incl. spiral pipetting tube, 3 mL 6.1562.050
6.10 Accessories 711 Liquino, Instructions for use 170 PTFE capillary tubing with 2xM6 thread, protected against light and buckling ∅ 0.3 mm insi
7. Index 711 Liquino, Instructions for use 171 7 Index Keys <> ... 65 <ALPHA> ...
7. Index 711 Liquino, Instructions for use 172 diluting ... 30; 100 diluting vol. ... 103
7. Index 711 Liquino, Instructions for use 173 optional accessories ... 167 options ... 71 organic s
2.5 Connections 711 Liquino, Instructions for use 12 711 Liquino — Titrino — 683 Pump — 730 Sample Changer Cable 6.2141.100 is used for the SM
7. Index 711 Liquino, Instructions for use 174 temperature measuring ... 29 temperature monitoring ... 79 temperature sensor
2. Installation 711 Liquino, Instructions for use 13 Switch sample changer to next sample*) >SEQ sequence st ep 14 ( CONT ROL command)
2.5 Connections 711 Liquino, Instructions for use 14 Example of a sample changer method: >st ar t sequence 1 CTL: Rm: INIT &g
2. Installation 711 Liquino, Instructions for use 15 731 Relay Box settings: Connect 772 Pump Unit to DC1 (or DC2) Output voltage: 18 or 24 volt
2.5 Connections 711 Liquino, Instructions for use 16 2.5.2 Serial connections A wide range of possible connections can be made at the RS232 int
2. Installation 711 Liquino, Instructions for use 17 2.5.3 Connecting a printer Printers with the following printer drivers can be connected:
2.5 Connections 711 Liquino, Instructions for use 18 Printer Cable RS232 settings Settings on printer Seiko DPU-414 6.2125.130 baud rate: 9
2. Installation 711 Liquino, Instructions for use 19 2.5.4 Connecting a balance The following balances can be connected to the RS232 output of
3.1 The Liquino concept 711 Liquino, Instructions for use 20 3 Introduction 3.1 The Liquino concept The 711 Liquino is designed so that, togeth
3 Introduction 711 Liquino, Instructions for use 21 3.2 Tutorial This tutorial will introduce you step by step to the operation of the Metrohm 7
3.2 Tutorial 711 Liquino, Instructions for use 22 3.2.2 Defining dosing units In order for the complete tubing system of a Dosino to be fully f
3 Introduction 711 Liquino, Instructions for use 23 di am. t ube2 2. 0 mm • The inner diameter of the tubing at Dosino port 2 can b
Metrohm AGCH-9101 HerisauSwitzerlandPhone +41 71 353 85 85Fax +41 71 353 89 [email protected] LiquinoManual8.711.1023 02.99 dm
3.2 Tutorial 711 Liquino, Instructions for use 24 3.2.3 Dosing and dispensing This is where you learn the simplest functions of the Liquino: ma
3 Introduction 711 Liquino, Instructions for use 25 • If you do not want the dosing speed to increase or you require a quicker increase in the d
3.2 Tutorial 711 Liquino, Instructions for use 26 MANUALMAN OFFD • Operating the <MAN. OFF / MANUAL> key switches manual operation off aga
3 Introduction 711 Liquino, Instructions for use 27 • If you want, you can configure your Liquino so that after each dispensing process the dosi
3.2 Tutorial 711 Liquino, Instructions for use 28 3.2.4 Changing the dosing unit A dosing device consists of a dosing unit (the green part) with
3 Introduction 711 Liquino, Instructions for use 29 3.2.5 Temperature measurement The 711 Liquino is equipped with a Pt100 temperature measuring
3.2 Tutorial 711 Liquino, Instructions for use 30 3.2.6 PIP – precise pipetting/diluting Example: Pipet 1mL and dilute with 9mL. Distilled wat
3 Introduction 711 Liquino, Instructions for use 31 <SELECT> <ENTER> <SELECT> <ENTER> <QUIT> dr
3.2 Tutorial 711 Liquino, Instructions for use 32 STARTZ <ENTER> • From now on when <START> is pressed the method will be run im-
3 Introduction 711 Liquino, Instructions for use 33 3.2.7 SMPL – fully automatic sampling The capabilities and possible uses of an SMPL method
TeachwareMetrohm AGCH-9101 [email protected] documentation is protected by copyright. All rights reserved.Although all the information
3.2 Tutorial 711 Liquino, Instructions for use 34 A new titration can now be carried out. The number of determinations and the number of times th
3 Introduction 711 Liquino, Instructions for use 35 The dosing units which correspond to the tubing used for the Dosinos must be defined. This pr
3.2 Tutorial 711 Liquino, Instructions for use 36 <ENTER> <ENTER> <ENTER> di l ut i on vol . 3 mL •
3 Introduction 711 Liquino, Instructions for use 37 <ENTER> <ENTER> <ENTER> <SELECT> <ENTER>
3.2 Tutorial 711 Liquino, Instructions for use 38 <ENTER> <ENTER> the value with <ENTER>. Di spl ay: >>aux.
3 Introduction 711 Liquino, Instructions for use 39 <ENTER> <QUIT> Di spl ay: par amet er s >SMPL pr esel ect i ons •
3.2 Tutorial 711 Liquino, Instructions for use 40 3.2.8 CONT – the automatic volumetric flask A typical daily task is the preparation of standa
3 Introduction 711 Liquino, Instructions for use 41 • Accept the entry with <ENTER>. Di spl ay: >CONT par amet er s meas. uni
3.2 Tutorial 711 Liquino, Instructions for use 42 A CONT method can be edited and stored in the method memory for each standard solution. Each
3 Introduction 711 Liquino, Instructions for use 43 3.2.9 XDOS – the electronic dropping funnel 120mL of solution are to be added in 5 hours. T
Table of content page 711 Liquino, Instructions for use 1 Overview 1 1.1 Application range __________________________________________1 1.2 Applicatio
3.2 Tutorial 711 Liquino, Instructions for use 44 J >XDOS par amet er s • <> is used to open the next submenu. Di spl ay: pa
3 Introduction 711 Liquino, Instructions for use 45 Various XDOS methods can be edited and stored in the method memory. Each of these methods c
3.2 Tutorial 711 Liquino, Instructions for use 46 3.2.10 XDOS – temperature-controlled addition Example: during the preparation for a nitration
3 Introduction 711 Liquino, Instructions for use 47 PARAMB <ENTER> <ENTER> • The <PARAM> key is used to access the paramete
3.2 Tutorial 711 Liquino, Instructions for use 48 <SELECT> <ENTER> <ENTER> <ENTER> <SELECT> <ENTER>
3 Introduction 711 Liquino, Instructions for use 49 <SELECT> <ENTER> <ENTER> 2x <QUIT> • In this menu
3.2 Tutorial 711 Liquino, Instructions for use 50 Running the method STARTZ • When <START> is now pressed the display shows Di spl ay:
3 Introduction 711 Liquino, Instructions for use 51 3.2.11 SEQ – complex procedures under control Example: the preparation of phenolazonaphthol
3.2 Tutorial 711 Liquino, Instructions for use 52 For the other two XDOS methods (diazotization and azo coupling) the pa-rameter lists are printe
3 Introduction 711 Liquino, Instructions for use 53 XDOS method for azo coupling 711 Li qui no 711. 0011 dat e 1998- 07- 08
Table of content page 711 Liquino, Instructions for use 4.4 XDOS mode - controlled dosing ____________________________ 74 4.4.1 XDOS mode parameters
3.2 Tutorial 711 Liquino, Instructions for use 54 SEQ method editing As the sub-methods have now been defined and stored, the SEQ method can now
3 Introduction 711 Liquino, Instructions for use 55 <SELECT> <ENTER> <SELECT> <ENTER> 2x <QUIT> comma
3.2 Tutorial 711 Liquino, Instructions for use 56 Running the method STARTZ • When <START> is now pressed the display shows Di spl ay:
3 Introduction 711 Liquino, Instructions for use 57 3.2.12 GLP – automatic validation For the automatic validation of a Dosino or the dosing un
3.2 Tutorial 711 Liquino, Instructions for use 58 Creating the GLP method *MODE-X DOS NRSELECTP • The GLP mode must first be activated. This se
3 Introduction 711 Liquino, Instructions for use 59 <ENTER> <ENTER> <SELECT> <ENTER> 2x <
3.2 Tutorial 711 Liquino, Instructions for use 60 Running the GLP method STARTZ • When <START> is now pressed the display shows Di spl
4. Detailed description Liquino 711, Instructions for use 61 4 Detailed description 4.1 The keyboard 711 Li qui no Cont rol l er DOS 1D
4.1 The keyboard Liquino 711, Instructions for use 62 4.1.1 Menu keys CONFIGA Opens the configuration menu. For details see page 69. PARAMB
4. Detailed description Liquino 711, Instructions for use 63 QUITNO Exits a submenu or confirms an error message. CLEARQ Completely deletes an e
1. Overview 711 Liquino, Instructions for use 1 1 Overview 1.1 Application range The Metrohm 711 Liquino is a very versatile laboratory instrum
4.1 The keyboard Liquino 711, Instructions for use 64 .PRINTW Prints out various reports. *MODE-X Selects an automatic run mode: • XDOS - contro
4. Detailed description Liquino 711, Instructions for use 65 4.2 Editing The instrument dialog system of the Liquino 711 is chiefly organized as
4.2 Editing Liquino 711, Instructions for use 66 4.2.3 Text input The text editor can be used wherever a text input is required. Numbers can be
4. Detailed description Liquino 711, Instructions for use 67 Example: >st or e met hod method: ******** First press the <CLEAR>
4.3 Configuration Liquino 711, Instructions for use 68 4.3 Configuration In order for the different modes and commands of the Liquino 711 to fun
4. Detailed description Liquino 711, Instructions for use 69 4.3.2 The configuration menu Please note: menu entries with a colon offer a selectio
4.3 Configuration Liquino 711, Instructions for use 70 >dosi ng uni t s Configuration of dosing units If dosing units with different lengths
4. Detailed description Liquino 711, Instructions for use 71 >>er r . l i mi t s 20mL- cyl . Submenu for 20 mL dosing units er r or max.
4.3 Configuration Liquino 711, Instructions for use 72 >keyboar d opt i ons l ock conf i gur at i on: of f Locks <CONFIG> key on o
4. Detailed description Liquino 711, Instructions for use 73 >show/ hi de modes hi de XDOS : no Hides XDOS mode yes no = the XDO
1.3 Instrument description 711 Liquino, Instructions for use 2 1.3 Instrument description 1.3.1 Front view Slot for memory card Always insert
4.4 XDOS mode - controlled dosing 711 Liquino, Instructions for use 74 4.4 XDOS mode - controlled dosing The XDOS mode (= Extended Dosing) of t
4. Detailed description 711 Liquino, Instructions for use 75 The following equation can be used to estimate whether the max. dosing rate will be
4.4 XDOS mode - controlled dosing 711 Liquino, Instructions for use 76 However, in order to ensure interruption-free dosing the following points
4. Detailed description 711 Liquino, Instructions for use 77 4.4.1 XDOS mode parameters PARAMB All settings in the parameter menu form a method
4.4 XDOS mode - controlled dosing 711 Liquino, Instructions for use 78 >XDOS par amet er fill rate ma x mL / mi n Filling rate of dosi
4. Detailed description 711 Liquino, Instructions for use 79 par amet er s >XDOS moni t or i ng Monitoring function settings Navigate with &l
4.4 XDOS mode - controlled dosing 711 Liquino, Instructions for use 80 >XDOS pr esel ect i ons >>XDOS pr ompt s Submenu for definition
4. Detailed description 711 Liquino, Instructions for use 81 4.4.2 Examples of reports - XDOS mode Temperature-controlled dosing with volume and
4.4 XDOS mode - controlled dosing 711 Liquino, Instructions for use 82 Curve Report 711 Liquino 0220 711.0011 date 1998-09-30
4. Detailed description 711 Liquino, Instructions for use 83 4.5 SEQ mode - complex processes The SEQ mode (sequence) of the Liquino allows comp
1. Overview 711 Liquino, Instructions for use 3 1.3.2 Rear view - Fire Hazard -For continued protection replaceonly with the same type andra
4.5 SEQ mode - complex processes 711 Liquino, Instructions for use 84 The process sequence is defined in the parameter menu and is part of the me
4. Detailed description 711 Liquino, Instructions for use 85 Auxiliary and communication commands Time-oriented auxiliary commands and communicati
4.5 SEQ mode - complex processes 711 Liquino, Instructions for use 86 RS232 connection: Data (=character strings) can be sent to connected instru
4. Detailed description 711 Liquino, Instructions for use 87 Input lines which are of no interest or for which no defined condi-tion can be predic
4.5 SEQ mode - complex processes 711 Liquino, Instructions for use 88 These status messages are only transmitted if first the corre-sponding stat
4. Detailed description 711 Liquino, Instructions for use 89 4.5.2 SEQ mode parameters PARAMB All settings in the parameter menu form a method
4.5 SEQ mode - complex processes 711 Liquino, Instructions for use 90 par amet er s >SEQ pr esel ect i ons Definition of requests at start o
4. Detailed description 711 Liquino, Instructions for use 91 4.5.3 Commands in a sequence METHOD >SEQ sequence st ep x command:
4.5 SEQ mode - complex processes 711 Liquino, Instructions for use 92 >SEQ sequence st ep x pat t er n *****--- Bit pattern of in
4. Detailed description 711 Liquino, Instructions for use 93 4.5.4 Example of a report - SEQ mode Sequences Parameter-Report 711 Li qui no